Product characteristics:
Active and passive testing;
Menu operation, simple and convenient;
The test history can be inquired and printed.
40000 data storage capacity, USB interface;
Printing integration;
project | Indicators and parameters | |
Measurement mode | Active blowing mode---Forensics, Continuous Passive Model---Investigation | |
sensor | Fuel Cell Electrochemical Alcohol Sensor | |
Input mode | 2.4Inch color touch screen+3Number of keys | |
Input content | License plate number, warning number, driver's license number, etc. | |
Printing mode | Integration with the mainframe | |
Active mode——Forensic Model | Measurement Range(BAC) | 0-550mg/100ml |
Blood Alcohol Concentration Range(BAC) | error | |
0-100mg/100ml | ±5mg/100ml | |
100-200mg/100ml | ±5% | |
>200mg/100ml | ±10% | |
Passive mode——Investigation Model | Measurement Range(BAC) | 0-100mg/100ml |
sensitivity | Very sensitive | |
Call the police | Red, yellow and green three-level alarm, alarm concentration can be set | |
Measuring time | Users can adjust the instrument themselves. | |
minimum | 1second | |
Longest | 10second | |
Working temperature range | -30℃~60℃ | |
Storage temperature range | -40℃~70℃ | |
Minimum blowing capacity | 1.0LOr set according to customer's needs | |
Blowpipe | One-way disposable or according to customer needs | |
Measuring time | less than10second | |
Clearing time | less than5second | |
Host Battery | 1300mAh/7.4VLithium ion rechargeable battery | |
Number of battery measurements | A full charge can work1000Sub-test | |
Display | 2.4In.26Multicolor, Full-color Dot Matrix LCD Touch Screen | |
storage | Date, time, license plate number, warning number, driver's license number, etc.(4Ten thousand records) | |
Product size | L170mm*W70mm*H30mm | |
Product weight | 125g | |
Calibration cycle | each6Once a month or as required |