细节展示放大镜特效

Belarus ATOMTEX Energy Spectrometer AT6102

  • No:AT6102
  • Quantity:

      Recommend

    Product Profile:
    AT6102, AT6102A and AT6102B are multi-functional portable energy spectrometers, which can measure gamma dose rate and detect and recognize various radionuclides without connecting computers.
    The instrument is mainly used for investigation of nuclear and radiation accidents, environmental radiation monitoring, customs and border control of illegal transportation of radioactive materials, etc.
    Inside the instrument, it can be stored.700 energy spectra, which can be used in extreme weather conditions, can also be connected to PC through USB data line.
    Gamma radiation can be measured by AT6102, AT6102A and AT6102B. Surface contamination of alpha and beta can be measured by connecting external probes BDPA-01 (a) and BDPB-01 (beta), and neutron radiation dose rate can be measured by connecting external probes BDKN-03.
    AT6102 differs from AT6102A and AT6102B in that two helium-3 proportional counters with moderators are built into the instrument for neutron radiation detection, neutron source searching and neutron counting rate measurement.
    The size, sensitivity and measurement range of detectors used in AT6102A and AT6102B are different. Specific technical parameters can be referred to.

    Product applications:
    Investigation of Nuclear Radiation Accidents
    Environmental Radiation Monitoring
    Nuclear Waste Control
    Border and Customs
    Nuclear Industry and Nuclear Power Station
    Explorable geology
    Radioactive drug production
    Nuclear medicine
    Nuclide studies

    Product characteristics:
    MeasurableAlpha, beta, gamma and neutron radiation
    Detection and location of radioactive sources, real-time identification of nuclides
    Multiple functions
    Strong and durable shape design
    Online measurement and professional analysis mode can be realized.
    Built-inGPS module
    By automatic continuous operationThe stability of LED is determined, and energy calibration is carried out regularly with reference material KCl.
    Temperature Compensation by Temperature Sensor
    Over Dose Threshold Alarm Function
    Ability to store700 energy spectra
    Able to measure in extreme environment on site
    adoptUSB interface is connected to computer
    External probe can be connected

    Product parameters:

    Energy Spectrometer Model

    АТ6102 (γ, n)

    AT6102A(γ)

    AT6102B(γ)

    Gamma detector

    NaI (TI),? 40x40 mm, GM counter

    NaI (TI),? 40x80 mm, GM counter

    Neutron Detector

    3He proportional neutron counter

    -

    -

    Gamma radiation energy range

    20 keV...3 MeV

    Energy range of neutron radiation

    0.025 eV...14 MeV

    -

    -

    Nuclide Recognition

    Medical nuclides, industrial nuclides, natural nuclides (nuclide Library of modifiable instruments)

    Resolving power

    7.5% FWHM
    (662 keV ,20°C)

    8% FWHM
    (662 keV ,20°C)

    Maximum Input Statistical Load

    ≥1.5·105 s-1

    Detection time (for 137Cs source, activity 50 kBq, 20 cm)

    ≤2s

    Detection time (252Cf, 10cm)

    ≤5 s

    Measurement range of gamma radiation dose rate

    0.01...300 μSv/h10μSv/h...100 mSv/h
    [NaI(Tl)][G-M]

    0.01...150 μSv/h
    10μSv/h...100 mSv/h
    [NaI(Tl)][G-M]

    Inherent error (dose rate)

    ±20%

    Gamma radiation sensitivity

    241Am

    6600 cps/μSv?h-1

    11600 cps/μSv?h-1

    137Cs

    850 cps/μSv?h-1

    1700 cps/μSv?h-1

    60Co

    430 cps/μSv?h-1

    840 cps/μSv?h-1

    Gamma radiation sensitivity (depending on energy)

    NaI(Tl)]
    50 keV—3 MeV

    ±20%

    [G M]
    60 keV—3 MeV

    ±25%

    Neutron Radiation Sensitivity

    Pu-Be

    0.28 Pulse & middot; cm2/n

    -

    -

    252Cf

    0.5 pulse middot; cm2/n

    -

    -

    A/D Conversion Channel Number

    1024

    Duration time

    No backlight

    ≥18h

    ≥25h

    Backlight

    ≥15h

    ≥17h

    Protection Level

    IP65

    Temperature range of operating environment

    -20°C...+50°C

    humidity

    ≤95%

    Connect to PC

    USB

    Dimensional weight

    230х115х212mm,

    2.5 kg

    230х115х177mm,

    1.9 kg

    230х115х177 mm,

    2.15kg

    Model of External Probe

    BDPA-01(α)

    BDPB-01(β)

    BDKN-03(n)

    detector

    ZnS(Ag),?60 mm

    Plastic scintillator,?60 mm

    3He proportional counter

    Energy range

    4-7MeV

    155keV-3.5MeV

    0.025 eV-14 MeV

    measuring range

    0.5-1·105
    P/(min & middot; cm2) radiation flux

    3-5-middot; 105p/(min-middot; cm2) radiation flux

    0.1 μSv/h -10 mSv/h
    dose rate

    sensitivity

    0.15cps/(part.·min-1·cm-)(239Pu)

    0.3cps/(part.·min-1·cm-2)(90Sr+90Y)

    0.355 cps/μSv?h-1
    (Pu-Be)

    Inherent error

    ±20%

    Protection Level

    IP54

    size

    87x205 mm, 0.55 kg

    87x205 mm, 0.65 kg

    314х220х263 mm, 8 kg