细节展示放大镜特效

Whole body gamma pollution monitor

  • No:WCM1000
  • Quantity:

      Recommend

    WCM1000 Whole Body Gamma Pollution Monitor
    WCM1000 Whole Body Gamma Contamination Monitor

    WCM1000 whole body gamma contamination monitor is installed at the health exit of the radiation control channel of nuclear power plant or related nuclear facility plant. It is used to monitor the gamma radiation contamination of all parts of the whole body of workers when they exit the radiation control zone, so as to detect the contamination in time and prevent the transfer of contamination. It is designed as a door cabinet structure with electric railings at the entrance and electric doors at the exit.
    Its main functions are:
    1. Inspection of the whole body pollution of staff and its level
    2. Find out the beginning of the accident by analyzing the data in time.
    3. Isolate control zones to avoid pollution diffusion in control zones
    4. Avoid unauthorized removal of radioactive sources from the control area

    Technical characteristics:
    1. Plastic scintillator detector
    2. Flat panel color display, showing the counting rate and total counting rate of each detector, indicating the polluted position.
    3. Speech warning during measurement, light and voice alarm when polluted
    4. The parameters of pollution measurement threshold and background measurement threshold of each detector can be adjusted by software.
    5. Provide calibration source and source frame, and calibrate instrument by software.
    6. Equipped with emergency channel button
    7. Personnel Approaching Detector and Personnel Position Sensor
    8. Timing background measurement and automatic background deduction

    Structural characteristics:
    External dimensions: 1200 mm * 850 mm * 2360 mm (width * thickness * height)
    Internal dimensions: 590 mm * 850 mm * 2060 mm (width * thickness * height)
    Weight: 1300 kg

    Technical Indicators:
    1. Detector size: trunk 500 mm * 500 mm * 50 mm (length * height * thickness)
    2. Head 500 mm * 300 mm * 50 mm (length * width * thickness)
    3. Foot 500 mm * 500 mm * 50 mm (length * width * thickness)
    4. Energy range: 50 keV-1.3 MeV, maximum error less than 50%.
    5. Sensitivity: greater than 10 cps/kBq (137 Cs, 30 cm from each detector center)
    6. Lower detection limit: 1.6 kBq (10 seconds measurement time, 99% confidence, 137Cs source in the middle of the monitor under 0.2 UG Gy/h environment background)
    7. Coefficient of variation: less than 10%
    8. Environmental Characteristics: Temperature 0-50 C, maximum humidity 90% (40 C)
    9. Power supply voltage: 220 V AC/50 Hz, UPS standby 2.5 hours
    10. Power Consumption: 220 V AC, 320 W